Cochlear SPrint Sound Processor Informazioni Techniche

Navigare online o scaricare Informazioni Techniche per no Cochlear SPrint Sound Processor. Cochlear SPrint Sound Processor Technical information Manuale Utente

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 78
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 0
Cochlear
Clinical Guidance Document
Vedere la pagina 0
1 2 3 4 5 6 ... 77 78

Sommario

Pagina 1 - Cochlear

Cochlear™ Clinical Guidance Document

Pagina 2

SECTION ONE | CLINICAL GUIDANCE81. Stop the stimulation and sweep all of the channels at 80% and then at C-level to confirm that sounds are comforta

Pagina 3

CLINICAL GUIDANCE | SECTION ONE94. The default streamlined programming channels 22, 16, 11, 6, and 1 are highlighted. In the Method box, click the

Pagina 4

SECTION ONE | CLINICAL GUIDANCE101.1.3 NRT/Objective Preset MethodThe NRT/objective preset method offsets the T- and C-level profiles from the objec

Pagina 5 - Getting started

CLINICAL GUIDANCE | SECTION ONE11where T-NRT fell in relation to the behavioural dynamic range. As a result of the large variability between recipie

Pagina 6 - T-level Assessment Procedures

SECTION ONE | CLINICAL GUIDANCE121.2 ADJUSTMENTS TO CHANNEL GAINSIn addition to the user-controlled microphone sensitivity that determines the overa

Pagina 7 - C-level Assessment Procedures

CLINICAL GUIDANCE | SECTION ONE13• Automatic Gain Control (AGC) – front-end amplifier that reduces the gain of high amplitude inputs to avoid distor

Pagina 8

SECTION ONE | CLINICAL GUIDANCE141.3.2 Supported Combinations of Implant System ComponentsThe supported combinations of implant system components an

Pagina 9 - 1.1.1 Behavioural Procedures

CLINICAL GUIDANCE | SECTION ONE151.4 ELECTRICAL STIMULATION AND MAP PARAMETERS• Stimulation Mode – describes the location of the indifferent electr

Pagina 10

SECTION ONE | CLINICAL GUIDANCE16• IIDR (Instantaneous Input Dynamic Range) – is the selected range of input signal at any instant in time that wil

Pagina 11

CLINICAL GUIDANCE | SECTION ONE171.5 STIMULATION MODESElectrical stimulation produces current flow between an active (stimulated) and indifferent (re

Pagina 13

SECTION ONE | CLINICAL GUIDANCE181.5.2 Bipolar (BP) StimulationIn BP stimulation, both the active and indifferent electrodes are inside the cochlea.

Pagina 14

CLINICAL GUIDANCE | SECTION ONE191.5.3 Variable and Pseudomonopolar (PSMP) StimulationVariable and pseudomonopolar stimulation modes combine differe

Pagina 15

SECTION ONE | CLINICAL GUIDANCE20Selecting the stimulation modeThe clinician sets the stimulation mode in the Custom Sound software on the “Open or

Pagina 16

CLINICAL GUIDANCE | SECTION ONE21• Out of Compliance – the condition where the maximum voltage available from the implant is not sufficient to gener

Pagina 17

SECTION ONE | CLINICAL GUIDANCE22Increasing the pulse width may result in the number of maxima being reduced automatically. The 3500 Hz per channel

Pagina 18 - Q varies

CLINICAL GUIDANCE | SECTION ONE23required programming changes. Note: Nucleus 24 system implant recipients may upgrade to the CP810 or Freedom proces

Pagina 19 - 1.5 STIMULATION MODES

SECTION ONE | CLINICAL GUIDANCE24The power level required is dependent on the type of MAP, its various parameters, and the individual recipient. Pow

Pagina 20 - Figure 3: Bipolar

CLINICAL GUIDANCE | SECTION ONE25If the coil is not on the implant when power optimization occurs, the power level is set to the previously saved le

Pagina 21

SECTION ONE | CLINICAL GUIDANCE26Before sending the recipient home, it is recommended that you determine the batteries that are best to use with the

Pagina 22 - 1.6 TELEMETRY MEASUREMENTS

CLINICAL GUIDANCE | SECTION ONE27The recipient does not hear this stimulation. For MAPs that are personalized for different listening environments,

Pagina 23 - 1.7.1 CI500 series, Freedom

1CLINICAL GUIDANCE | TABLE OF CONTENTSSECTION 1 : PROGRAMMING PRINCIPLES1.1 STREAMLINED PROGRAMMING METHODS 61.1.1 Behavioural Procedures 71.1.2

Pagina 24 - 24 System

SECTION ONE | CLINICAL GUIDANCE28Table 1: The differences between SPEAK, CIS/CIS (RE) and ACE/ACE (RE)StrategyNo. of stimulation sitesStimulation ra

Pagina 25 - 1.8 POWER OPTIMIZATION

CLINICAL GUIDANCE | SECTION ONE291.9.2 SPEAKSPEAK (Spectral Peak) focuses on the spectral (frequency) properties of sound. It takes advantage of the

Pagina 26

SECTION ONE | CLINICAL GUIDANCE30The selected channels receive sequential stimulation in tonotopic order from high to low frequency. SPEAK uses a st

Pagina 27

CLINICAL GUIDANCE | SECTION ONE311.9.3 CIS/CIS (RE)CIS (Continuous Interleaved Sampling) focuses on the temporal (timing) properties of sound. It us

Pagina 28

SECTION ONE | CLINICAL GUIDANCE32Figure 8 illustrates the electrical stimulation pattern of the English word ‘choice’ using a 6-channel CIS strategy

Pagina 29 - 1.9 SOUND CODING STRATEGIES

CLINICAL GUIDANCE | SECTION ONE33When programming a Nucleus® 24 implant and the SPrint processor with 20 active channels, the default is Table 8 (18

Pagina 30 - /ACE (RE)

SECTION ONE | CLINICAL GUIDANCE341.11 MICROPHONE SENSITIVITY AND VOLUME SETTING1.11.1 Microphone SensitivityThe setting of the microphone sensitivit

Pagina 31 - 1.9.2 SPEAK

CLINICAL GUIDANCE | SECTION ONE35softer speech sounds. In both cases, C-levels should be re-assessed.The Sensitivity Mode may be set to ‘Manual’ or

Pagina 32

SECTION ONE | CLINICAL GUIDANCE36With the CP810 Sound processor, the clinician can lock out the ability to change both sensitivity and volume via th

Pagina 33 - 1.9.3 CIS/CIS (RE)

CLINICAL GUIDANCE | SECTION ONE37The channel gains adapt so as to optimize the loudness of the signal into the recipient’s dynamic range. For low in

Pagina 34

2SECTION ONE | CLINICAL GUIDANCEThe primary aim of programming a sound processor is to customize the device so that the cochlear implant provides co

Pagina 35

SECTION ONE | CLINICAL GUIDANCE38Whisper may not provide significant benefit to all recipients in all environments. Past studies2,3 with Whisper have

Pagina 36 - 1.11.1 Microphone Sensitivity

CLINICAL GUIDANCE | SECTION ONE39Whisper is used alone and in combination for the SmartSound Everyday environment and also in combination for the Sm

Pagina 37 - 1.11.2 Volume Setting

SECTION ONE | CLINICAL GUIDANCE40Focus with the Freedom Sound ProcessorThe Freedom sound processor implements the beam algorithm using one dual port

Pagina 38 - 1.12.1 ADRO

CLINICAL GUIDANCE | SECTION ONE41The polar plots in Figure 14 compare the directional sensitivity of Beam (light blue line) and a fixed directional m

Pagina 39 - 1.12.2 Whisper

SECTION ONE | CLINICAL GUIDANCE42The effects of Beam were evaluated in a double-blind trial with five adult Nucleus® CI users4. Subjects were tested

Pagina 40 - Whisper boosts soft

CLINICAL GUIDANCE | SECTION ONE43Group mean results for CUNY sentences at 65dB SPL are shown in Figure 16, and indicate a statistically significant m

Pagina 41 - 1.12.3 Focus (using Beam

SECTION ONE | CLINICAL GUIDANCE44equally, so a null position is selected to give the best overall result. With zoom, this null direction is fixed (Fi

Pagina 42 - Figure 13 : A block

CLINICAL GUIDANCE | SECTION ONE45References1. CP810/CR110 Validation, Cochlear internal data Aug 2009.1.12.5 Autosensitivity™ (ASC)The function of t

Pagina 43

SECTION ONE | CLINICAL GUIDANCE46Recommendations for changing Autosensitivity (ASC) breakpointIncreasing the Autosensitivity breakpoint will make it

Pagina 44

CLINICAL GUIDANCE | SECTION ONE47Environment OptionsEVERYDAYThis environment is used in typical listening situations. A clinician will typically pro

Pagina 45 - 1.12.4 Zoom

3CLINICAL GUIDANCE | SECTION ONESECTION 1 : PROGRAMMING PRINCIPLESGetting startedAt the initial activation, the clinician uses Custom Sound™ to prog

Pagina 46 - Processor

SECTION ONE | CLINICAL GUIDANCE48FOCUSThis environment is used when there is significant background noise, but the focus is on hearing what one perso

Pagina 47 - 1.12.5 Autosensitivity

CLINICAL GUIDANCE | SECTION ONE49When you convert a Freedom MAP to a Nucleus 24 MAP, input processing options that are not supported on Nucleus 24 s

Pagina 48 - ENVIRONMENTS

SECTION ONE | CLINICAL GUIDANCE50• Alternating stimulus polarity will effectively reduce the stimulus artefact. However, there may be a disadvantag

Pagina 49 - EVERYDAY

CLINICAL GUIDANCE | SECTION ONE51Absence of ESRT does not mean the implant is not working or the recipient is not hearing. There can be other causes

Pagina 50

SECTION ONE | CLINICAL GUIDANCE52Accessory Adaptor. The Freedom Accessory Adaptor is needed to connect the processor end of the ESPrit FM cable into

Pagina 51 - 1.14 OBJECTIVE MEASUREMENTS

CLINICAL GUIDANCE | SECTION ONE53Parameter Recommendation RationaleAuto Telecoil Off. (Note: for neckloop receivers, please activate the telecoil)Co

Pagina 52

SECTION ONE | CLINICAL GUIDANCE54Many factors impact the battery life of a sound processor including the use of some accessories. It is important to

Pagina 53 - 1.15.1 CP810 Sound Processors

CLINICAL GUIDANCE | SECTION TWO55SECTION TWO | CLINICAL GUIDANCESECTION 2: PROGRAMMING RECOMMENDATIONS FOR SPECIALTY DEVICES2.1 SPECIALTY DEVICES

Pagina 54

SECTION TWO | CLINICAL GUIDANCE562.1.1.1 Setting Up the ABI EquipmentBefore switch-on• Since tuning the Nucleus 24 ABI is time-consuming, consider

Pagina 55

CLINICAL GUIDANCE | SECTION TWO57ECG monitoring equipmentCaution: During stimulation the recipient must be connected to ECG monitoring equipment wit

Pagina 56

SECTION ONE | CLINICAL GUIDANCE4• Current Level (CL) – the amount of electrical current delivered to the implant recipient expressed in clinical pr

Pagina 57 - SPECIALTY DEVICES

SECTION TWO | CLINICAL GUIDANCE587. Position the transmitter coil so it is just below the retainer disk. The transmitter coil will then be located d

Pagina 58 - Nucleus 24 ABI

CLINICAL GUIDANCE | SECTION TWO59Performing PsychophysicsNote: It is important that you perform psychophysics on an electrode-by-electrode basis. In

Pagina 59 - 2.1.1.2 Programming the ABI

SECTION TWO | CLINICAL GUIDANCE60Scenario OneIn the best case scenario, all channels would have both T-levels and C-levels below any side-effect thr

Pagina 60 - Loudness 1 – 10

CLINICAL GUIDANCE | SECTION TWO61Scenario ThreeIn a typical case, there is usually a group of channels which has only side effects, another which ha

Pagina 61

SECTION TWO | CLINICAL GUIDANCE62Table 1: Loudness/sensation strength scaleAuditory Sensation Scale Side EffectStop 10 Stop Too loud 9 Uncomfortable

Pagina 62

CLINICAL GUIDANCE | SECTION TWO63What to do if you reach the safe limit levelThe levels listed in the table below are considered to be conservative

Pagina 63

SECTION TWO | CLINICAL GUIDANCE64Table 3: Maximum safe current and corresponding current levels for various stimulus pulse widths for the Nucleus AB

Pagina 64

CLINICAL GUIDANCE | SECTION TWO65For supplementary programming guidelines for the Nucleus Freedom Processor for Nucleus 24 ABI (ABI24M) using Custom

Pagina 65

SECTION TWO | CLINICAL GUIDANCE662.1.1.4 Fine Tuning the ABIBalancing/sweepingBefore carrying out any pitch perception test, it is important to bala

Pagina 66 - Figure 8 : T-levels

CLINICAL GUIDANCE | SECTION TWO67Give clear instructions to the recipient before conducting the pitch-ranking procedure to ensure they understand th

Pagina 67 - 2.1.1.3 Managing Side Effects

CLINICAL GUIDANCE | SECTION ONE5The remaining T-levels are automatically interpolated. If you wish to measure other channels, simply double-click on

Pagina 68 - 2.1.1.4 Fine Tuning the ABI

SECTION TWO | CLINICAL GUIDANCE6811.It is strongly advised that after every few channels have been added to the ordered list, to sweep all the order

Pagina 69

CLINICAL GUIDANCE | SECTION TWO69a better subjective impression between MAPs. Once you select a favourable MAP, encourage the recipient to use the p

Pagina 70 - ABI 24M pitch

SECTION TWO | CLINICAL GUIDANCE702.1.1.7 ABI WorksheetsFigure 10 : Nucleus® 24 ABI MAPping worksheetFigure 11 : MAPping an ABI 24 recipient

Pagina 71 - 2.1.1.6 Rehabilitation

CLINICAL GUIDANCE | SECTION TWO71Note: It is recommended to maximize perceptual abilities wherever possible. This is because ABI performance is gene

Pagina 72 - 2.1.1.7 ABI Worksheets

SECTION TWO | CLINICAL GUIDANCE722.1.2 THE NUCLEUS® 24 DOUBLE ARRAYThe Nucleus® 24 Double Array (CI 11+11+2M) cochlear implant is based on the Nucle

Pagina 73

CLINICAL GUIDANCE | SECTION TWO73Prior to programming, check with the surgery report to confirm which electrode array is in which turn of the cochlea

Pagina 74 - 24 DOUBLE ARRAY

SECTION TWO | CLINICAL GUIDANCE74To ascertain electrode location (base or apex):1. Select one active electrode mid-way along each array (e.g. 6 and

Pagina 75

CLINICAL GUIDANCE | NOTES75

Pagina 76

As the global leader in hearing solutions, Cochlear is dedicated to bringing the gift of sound to people all over the world. With our hearing solution

Pagina 77 - CLINICAL GUIDANCE

SECTION ONE | CLINICAL GUIDANCE61.1 STREAMLINED PROGRAMMING METHODSCustom Sound includes three streamlined programming methods that make programming

Pagina 78

CLINICAL GUIDANCE | SECTION ONE71.1.1 Behavioural Procedures1. Measure impedances by performing an implant test.2. In the Open or Create MAP scre

Commenti su questo manuale

Nessun commento